Use of Beyond-Linearity Quality Control Sera for Routine Internal Quality Control of Glucose and Creatinine

Use of beyond-linearity quality control sera for routine internal quality control of Glucose and Creatinine

Authors

  • Nilam Karbhari
  • Dipti Karbhari
  • Shailesh Patel

Keywords:

Linearity, Quality Control, LJ Chart, Creatinine, Glucose

Abstract

Introduction: While linearity of assay is tested during validation, there is no day to day check on changes thereof, although change in linearity will affect results around the linearity limits. Objectives: The study was performed to see if introduction of such beyond linearity QC program is useful to the laboratory or not. Methods: Serum control pool of beyond linearity range was prepared and analysed along with the routine quality controls and the incidences of rejections of routine internal controls and beyond linearity controls were compared. Results: There was general correlation in rejection incidences between routine QC and Beyond-linearity QC for both 1(3S) and 2(2S) QC rules. Interpretation and conclusion: The study shows that, there is no major increase in error detection rate on introduction of beyond-linearity QC. However, there was increase in warnings 1(2S) signs with beyond linearity QC. [Nilam K NJIRM 2017; 8(1): 77-81]

References

1. Carl AB, Edward RA, David EB, et al. Quality Management. In: Klee GG, Westgard JO, editors. Tietz Textbook of Clinical Chemistry and Molecular Diagnostics, 5th Edi. New Delhi: Elsevier, a division of reed Elsevier India Private Limited; 2012.p.163-204
2. Ferrero CA, Carobene A, Ceriotti F, Modenese A, Arcelloni C. Behavior of Frozen Serum Pools and Lyophilized Sera in an External Quality-Assessment Scheme. Clin. Chem. 1995;41(4):575-80.
3. Bowers GN, Burnett RW, Mccomb RB.Preparation and Use of Human Serum Control Materials for Monitoring Precision in Clinical Chemistry. Clin. Chem. 1975;21(12):1830-36.
4. NCCLS. Statistical Quality Control for Quantitative Measurements: Principles and Definitions; Approved Guideline-Second Edition. NCCLS Document C24-A2 (ISBN1-56238-371-X). NCCLS, 940 West Valley Road, Suite 1400, Wayne, Pennsylvania, 19087 USA, 1999.
5. http://www.randox.com [internet]. 55, Diamond Road, Crumlin, County Antrim, BT294QY, United Kingdom. [Updated on: 05/12/2013 Cited on:2016-12]. Available from: http://inserts.randox.com/view_file.php?aku=c3Vi
bWl0PXZpZXcmaWQ9MjU3MCZtaW1ldHlwZT1hcHBsaWNhdGlvbiUyRnBkZg==
6. http://www.randox.com [internet]. 55, Diamond Road, Crumlin, County Antrim, BT294QY, United Kingdom. [Updated on: 21/05/2013 Cited on:2016-12]. Available from: http://inserts.randox.com/view_file.php?aku=c3VibWl0PXZpZXcmaWQ9MjUzNCZtaW1ldHlwZT1hcHBsaWNhdGlvbiUyRnBkZg==
7. http://www.randox.com [internet]. 55, Diamond Road, Crumlin, County Antrim, BT294QY, United Kingdom. [Cited on:2016-12]. Available from: http://inserts.randox.com/view_file.php?aku=c3VibWl0PXZpZXcmaWQ9Mjk3MiZtaW1ldHlwZT1hcHBsaWNhdGlvbiUyRnBkZg==
8. http://www.randox.com [internet]. 55, Diamond Road, Crumlin, County Antrim, BT294QY, United Kingdom. [Cited on:2016-12]. Available from:http://inserts.randox.com/view_file.php?aku=c3VibWl0PXZpZXcmaWQ9Mjk3MCZtaW1ldHlwZT1hcHBsaWNhdGlvbiUyRnBkZg==
9. http://www.randox.com [internet]. 55, Diamond Road, Crumlin, County Antrim, BT294QY, United Kingdom. [Cited on:2016-12]. Available from:http://inserts.randox.com/view_file.php?aku=c3VibWl0PXZpZXcmaWQ9MjI1NCZtaW1ldHlwZT1hcHBsaWNhdGlvbiUyRnBkZg==
10. http://www.bio-rad.com [Internet]. Bio-Rad Laboratories IndiaPvt.Ltd., "Bio-RadHouse" 86-87, UdyogVihar, Phase-IV Gurgaon- 122015 Haryana, India. [Cited on:2016-12] Available from: http://fs.classic.qcnet.com/ShowPDF.aspx?fName=BI+45660+MA-2015+08+31&rptType=pi
11. NCCLS. Statistical Quality Control for Quantitative Measurements: Principles and Definitions; Approved Guideline-Second Edition. NCCLS Document EP06-A. NCCLS, 940 West Valley Road, Suite 1400, Wayne, Pennsylvania, 19087 USA, 1999.

Downloads

Published

2018-01-31

How to Cite

Karbhari, N., Karbhari, D., & Patel, S. (2018). Use of Beyond-Linearity Quality Control Sera for Routine Internal Quality Control of Glucose and Creatinine: Use of beyond-linearity quality control sera for routine internal quality control of Glucose and Creatinine. National Journal of Integrated Research in Medicine, 8(1), 77–81. Retrieved from http://nicpd.ac.in/ojs-/index.php/njirm/article/view/1175

Issue

Section

Original Articles